Agilent Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is available in single quadrupole and triple quadrupole configurations. Our single quadrupole ICP-MS address the range of typical ICP-MS measurements from routine, regulated analysis to advanced research and demanding applications. Our triple quadrupole ICP-MS will give you the confidence of knowing you can resolve problematic interferences – even in your most difficult samples – delivering results you know you can trust.

Catalog No. Product Net Price
CP913777 Sample inlet tubing for inert Mira Mist nebulizer used with ICP-OES/ICP-MS. Fitted with 0.5 m of 0.018 inch id x 0.044 inch od tubing with threaded fitting to connect directly to nebulizer body. 137 
Add to Cart
  • Inner Diameter (ID)
  • 0.46 mm
  • Length
  • 0.5 m
  • UNSPSC Code
  • 41115411
  • Outer Diameter (OD)
  • 1.11 mm
  • Where Used
  • Nebulizer
G1820-65138 Nebulizer, concentric, quartz, recommended for organic and high purity analysis. Quartz concentric nebulizer recommended for trace analysis of high purity samples. Cleaner than borosilicate glass nebulizers used as standard. Suitable for most sample type 1351 
Add to Cart
  • UNSPSC Code
  • 41115411
  • Autosampler Compatibility
  • I-AS Other SPS 3 SPS 4
  • Chemical Compatibility
  • Aqueous Organic solvents
  • Material
  • Quartz
  • Nebulizer Sample Particle Size Tolerance
  • 40 µm
  • Notes
  • Recommended for trace analysis of clean samples. Lower contamination level than Pyrex. Clean using 5 % HNO3 (v/v) bath; do not use ultrasonic bath. 3 (v/v) bath; do not use ultrasonic bath. -->
  • Self Aspirating
  • Yes, volume not limited
  • Type
  • Concentric
G1833-65583 Nebulizer gas line connector for Microflow PFA concentric nebulizer. Union connector for nebulizer (a.k.a. carrier) argon gas line 30 
Add to Cart
  • UNSPSC Code
  • 41115411
  • Material
  • PFA
  • Notes
  • Connector for MicroFlow nebulizer
  • Where Used
  • Nebulizer
G1960-20033 Microflow nebulizer needle 183 
Add to Cart
  • UNSPSC Code
  • 41116107
G3161-80001 Nebulizer, Mira Mist, PEEK, for high TDS applications, suitable for all Agilent ICP-MS series. Enhanced parallel path nebulizer gives high salts and TDS capability. Not suitable for HF, high concentrations of acids and some organic solvents. Does not sel 1130 
Add to Cart
  • UNSPSC Code
  • 41115411
  • Autosampler Compatibility
  • I-AS Other SPS 3 SPS 4
  • Chemical Compatibility
  • Aqueous
  • Gas Connector
  • Threaded connector
  • Material
  • PEEK
  • Name
  • MiraMist
  • Nebulizer Sample Particle Size Tolerance
  • 150 µm
  • Notes
  • Agilent 7500/7700 and 8800 series ICP-MS
  • Sample Connection
  • Threaded connector
  • Sample Uptake Rate
  • 2500
  • Self Aspirating
  • No
  • Type
  • Parallel path
G3266-80004 Nebulizer, MicroMist, U-series with UniFit sample connector, standard with Agilent 7800/7850/7900 Series ICP-MS. Borosilicate glass concentric nebulizer used as standard on Agilent x-lens ICP-MS instruments. Suitable for most sample types except those con 789 
Add to Cart
  • UNSPSC Code
  • 41113000
  • Autosampler Compatibility
  • I-AS Other SPS 3 SPS 4
  • Chemical Compatibility
  • Aqueous Organic solvents
  • Gas Connector
  • Ezylok
  • Material
  • Borosilicate glass
  • Name
  • MicroMist
  • Nebulizer Sample Particle Size Tolerance
  • 40 µm
  • Notes
  • Standard on 78/79/8800Suitable for most sample types except those containingHF or suspended particulates. Suitable for most sample types except those containingHF or suspended particulates. -->
  • Sample Connection
  • U-Series
  • Sample Uptake Rate
  • 400
  • Self Aspirating
  • Yes, volume not limited
  • TDS Handling Capability
  • 15 % TDS
  • Type
  • Concentric
G3266-80005 Nebulizer, MicroMist, U-series with UniFit sample connector and threaded Direct Connection gas line, standard with Agilent 8900 Series ICP-MS. Borosilicate glass concentric nebulizer used as standard on Agilent x-lens ICP-MS instruments. Suitable for most 736 
Add to Cart
  • UNSPSC Code
  • 41113000
  • Autosampler Compatibility
  • I-AS Other SPS 3 SPS 4
  • Chemical Compatibility
  • Aqueous Organic solvents
  • Gas Connector
  • Threaded connector
  • Material
  • Borosilicate glass
  • Name
  • MicroMist
  • Nebulizer Sample Particle Size Tolerance
  • 40 µm
  • Notes
  • Suits Agilent 8900 Series ICP-MS
  • Sample Connection
  • U-Series
  • Sample Uptake Rate
  • 400
  • Self Aspirating
  • Yes, volume not limited
  • TDS Handling Capability
  • 15 % TDS
  • Type
  • Concentric
G3266-80011 Sample uptake tube and connector, EzyFit, 0.5 mm id, 700 mm capillary, for MicroMist concentric nebulizer, 10/pk. Standard on 7700 and 8800 ICP-MS, suitable for all Agilent ICP-MS using the EzyFit MicroMist nebulizer 245 
Add to Cart
  • Inner Diameter (ID)
  • 0.5 mm
  • Length
  • 700 mm
  • UNSPSC Code
  • 41115411
  • Notes
  • For MicroMist nebulizer with EzyFit sample connector, 0.5 mm id, 700 mm
  • Where Used
  • Nebulizer
G3266-80012 Sample uptake tube and connector, UniFit, 0.5 mm id, with 0.5 mm id x 700 mm capillary, for U-Series MicroMist concentric nebulizer, 10/pk. Standard on 7800, 7850 and 7900 ICP-MS, suitable for all Agilent ICP-MS using the U-series MicroMist nebulizer 236 
Add to Cart
  • Inner Diameter (ID)
  • 0.5 mm
  • Length
  • 700 mm
  • UNSPSC Code
  • 41113000
  • Notes
  • For U-Series MicroMist nebulizer with Unifitsample connector sample connector -->
  • Where Used
  • Nebulizer
G3266-80015 EzyLok gas connector for carrier gas, suits most concentric nebulizers used with ICP-OES/MS. Only Ezylok connector supplied, suits 4 mm od gas inlet tubing 33 
Add to Cart
  • UNSPSC Code
  • 41115411
  • Where Used
  • Nebulizer

Overview

ICP-MS system operates with an ICP-MS Nebulizer, in which a high velocity Argon gas stream meets a liquid stream causing filmwise breakup of the solution into a fine aerosol. The aerosol particle size should ideally be sub-10µm with a narrow particle size distribution to achieve efficient atomization and ionization. Agilent offers a comprehensive range of ICP-MS nebulizers and supplies to solve any ICP-MS analysis challenge.

The portfolio includes standard glass concentric nebulizer (MicroMist), quartz and inert PFA nebulizers (MicroFlow and C-Flow), as well as robust Mira Mist PEEK or PTFE for high TDS samples and low-flow self-aspirating or total consumption nebulizers for your clinical or limited volume samples. ICP-MS nebulizer support is provided with a full range of accessories for all models including nebulizer cleaning tools.


The ICP torch is used to form, shape, and sustain an argon plasma which atomizes and ionizes the sample aerosol. Agilent ICP-MS Torches with Shield Torch System (STS) provide precise ion energy control, ensuring high sensitivity and stability, even in cool plasma mode for effective interference removal (ArO+).

Our portfolio is comprised of a wide range of ICP torches for your specific ICP-MS configuration. One-piece quartz torches offer the best stability and ease of use. The large 2.5mm injector on the standard torch gives maximum robustness and lowest matrix interferences for the best plasma stability. Single-piece torches with narrow bore injectors (1.0mm or 1.5 mm) are designed for use with organic solvents, to reduce plasma loading and improve plasma stability. Optional semi-demountable quartz torches with removable platinum or sapphire injectors are also available for use with the PFA inert sample introduction kit.