Altitude
2000 m (6562 ft) maximum
-
Analog Output Functional Mode
Linear, PID
-
Communication Capabilities
Profibus DPV1 card
-
Compatible Instruments
Ammonium: Amtax sc, AN-ISE sc, A-ISE scNitrate: Nitratax sc, N-ISE sc, Nitratax clear/eco/plus scPhosphate: Phosphax sc, Phosphax sc LROrganics: Uvas scpH/ORP: pHD sc, 1200-S sc, 8362scConductivity: 3700sc, 3798-S sc, 3400scOzone: 9185scDissolved Oxygen: LDO 2 scChlorine: CL10sc, CL17sc, 9184scChlorine dioxide: 9187scSuspended Solids: Solitax sc, TSS scSludge Level: Sonatax scTurbidity: TU5300sc/TU5400sc, Ultraturb sc, SS7 sc
-
Compatible Network Technologies
GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.)External CDMA (e.g. Verizon)
-
Compliance Certifications
CE approved (with all sensor types). Listed for use in general locations to UL and CSA safety standards by ETL (with all sensor types).Certain AC mains powered models are listed for use in general safety locations to UL and CSA safety standards by Underwriters Laboratories (with all sensor types).
-
Conduit Openings
½" NPT conduit
-
Dimensions
½ DIN - 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.)
-
Display
None
-
Enclosure Rating
NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
-
Installation Category
Category II
-
Material Enclosures
Polycarbonate, aluminum (powder coated), stainless steel
-
Measurements
One connection to SC controller
-
Mounting
Wall, Pole, or Panel Mounting
-
Network Connectivity
LAN version: Two Ethernet connectors (10/100 Mbps), switch function, M12 female D-coding connector.Cellular version: SIM card holder (3G/4G).The type of connection is dependent on controller configuration.
-
Operating Temperature Range
LAN version: -20 - 60 °C (-4 - 140 °F)Cellular version: -20 - 55 °C (-4 - 131 °F)
-
Pollution Degree
2
-
Power Requirements (Voltage)
100 - 240 VAC
-
Protection Class
I, connected to protective earth
-
Relays
High voltage:Two relays (SPDT)Maximum switching voltage: 250 VACMaximum switching current: 5 A (resistive only load)Maximum switching power: 1250 VA, 125 W (resistive only load)Wire gauge: 0.75 to 2.5 mm2 (18 to 12 AWG)Low voltage:Four relays on each low voltage relay module
-
Storage Conditions
-20 - 70 °C (-4 - 158 °F), 0 - 95% relative humidity, non-condensing
-
USB Port
Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor.
-
Weight
3.7 lb (controller only, w/o modules)
-
What's included?
SC4200c Claros Connector, LAN, User Manual, and Power Cable.Remote Service Connector Cable needs to be ordered separately.