+852-3069 6950
sales@gugent.com.hk
Catalog No. HACH9582.99.70P4
Part No. 9582.99.70P4
Supplier HACH Company
Polymetron 9582 Dissolved Oxygen System, 24 V DC
Net Price
Special offer
Availability

6 weeks

PROPERTIES
  • Cable Length
    10 m (33 ft)
  • Calibration Method
    Zero: Electrically or with oxygen free water, Slope: in air or against a laboratory measurement
  • Communication Capabilities
    Standard (2x 4-20 mA outputs)
  • Conduit Openings
    1/2" NPT Conduit
  • Display
    Graphic dot matrix LCD with LED backlighting, transreflective
  • Enclosure Rating
    IP66 / NEMA 4X
  • Maintenance Interval
    Membrane Lifetime: 6 months depending on sample
  • Material
    Electrodes: Gold cathode, Silver anode; Flow-through cell: Stainless Steel 316L; Membrane: PFA, Sensor Body: Noryl
  • Material Enclosures
    PolycarbonateAluminium (powder coated)Stainless Steel
  • Measurements
    0 to 2000 ppb (0-2 ppm)
  • Operating Temperature Range
    -20 - 60 °C at 0 - 95% RH (non-condensing)
  • Power Requirements (Voltage)
    24 V DC
  • Range
    0 - 2000 ppb (0-2 ppm)
  • Relays
    Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
  • Repeatability
    ± 0.5 ppb or ±5%, whichever is greater
  • Storage Conditions
    -20 °C to 70 °C
  • Warranty
    12 months
  • Weight
    3.2 kg
  • What's included?
    Controller, Mounting Hardware, Oxygen sensor, Temperature Sensor, Cable, Flow-Chamber, Basic User Manual
DESCRIPTION
An integral part of the most complete water analytics system for the Power industry. Hach provides a broad range of product options designed to work together into flexible solutions to meet your unique needs. Hach’s comprehensive approach saves you time on design, installation, training, maintenance, and operation.
CUSTOMERS ALSO VIEWED
RELATED CATEGORIES