+852-3069 6950
sales@gugent.com.hk
Catalog No. HACH9582.99.75P4
Part No. 9582.99.75P4
Supplier HACH Company
Polymetron 9582 Dissolved Oxygen System with HART Communications, 24 V DC
Net Price
Special offer
Availability

6 weeks

PROPERTIES
  • Cable Length
    10 m (33 ft)
  • Calibration Method
    Zero: Electrically or with oxygen free water, Slope: in air or against a laboratory measurement
  • Communication Capabilities
    Hart
  • Conduit Openings
    1/2" NPT Conduit
  • Display
    Graphic dot matrix LCD with LED backlighting, transreflective
  • Enclosure Rating
    IP66 / NEMA 4X
  • Maintenance Interval
    Membrane Lifetime: 6 months depending on sample
  • Material
    Electrodes: Gold cathode, Silver anode; Flow-through cell: Stainless Steel 316L; Membrane: PFA, Sensor Body: Noryl
  • Material Enclosures
    PolycarbonateAluminium (powder coated)Stainless Steel
  • Measurements
    0 to 2000 ppb (0-2 ppm)
  • Operating Temperature Range
    -20 - 60 °C at 0 - 95% RH (non-condensing)
  • Power Requirements (Voltage)
    24 V DC
  • Range
    0 - 2000 ppb (0-2 ppm)
  • Relays
    Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
  • Repeatability
    ± 0.5 ppb or ±5%, whichever is greater
  • Storage Conditions
    -20 °C to 70 °C
  • Weight
    3.2 kg
  • What's included?
    Controller, Mounting Hardware, Oxygen sensor, Temperature Sensor, Cable, Flow-Chamber, Basic User Manual
DESCRIPTION
An integral part of the most complete water analytics system for the Power industry. Hach provides a broad range of product options designed to work together into flexible solutions to meet your unique needs. Hach’s comprehensive approach saves you time on design, installation, training, maintenance, and operation.