+852-3069 6950
sales@gugent.com.hk
Catalog No. CYT18110303
Part No. 18110303
Supplier Cytiva
BPG 100/950 Column
Net Price
Special offer
Availability

4 weeks

PROPERTIES
  • Bed volume
    0.3-2 l
  • Dimensions
    480 x 480 x 1270 mm
  • Adapter Movement
    Manual
  • Autoclavable
    121°C for 30 min, except for the bed support
  • Bed height
    1-26 cm
  • Bed Height Packing With Extension Max.
    35 cm
  • Bed Height Running With Extension Max.
    45 cm
  • Bed support
    Net
  • Bed Volume Packing With Extension Max.
    2.7 l
  • Bed Volume Running With Extension Max.
    3.5 l
  • Column Cross Section Area
    78.5 cm²
  • Compliance
    The product fulfills valid directives and standards when used within the conditions specified in the user manual. The product must also be used in the same state as it was delivered from Cytiva and connected only to other CE labeled Cytiva modules or other products as recommended. For regulatory details please see declaration of conformity.
  • Connections
    25 mm TC
  • Footprint
    480 x 480 mm
  • Inlet Dimension i.d.
    4
  • Material (bed support)
    Polyamid (PA), Polyetheretherketone (PEEK), Polypropylene (PP)
  • Material (column tube)
    Glass Borosilicate 3.3
  • Material (Lid)
    316L Stainless Steel (SS)
  • Material (Seal)
    Ethylene propylene diene (EPDM), Fluoroethenepropene (FEP)
  • Material (Stand)
    316L Stainless Steel (SS)
  • Operating pressure max.
    8 bar (116.03 psi)
  • Operating Temperature
    4°C-40°C
  • Order Information
    Stand not included for column sizes 100-200
  • Outer Height
    1270
  • Porosity
    10 µm, 12 µm, 23 µm and 54 µm
  • Surface Finish, Stainless Steel Internal Wetted
    Glass Ra < 0.4 μm, SS UNS S32205, SS UNS S31803, SS ASTM 316L Ra 0.5 μm
  • Surface Finish Stainless Steel Non-Wetted Parts
    Ra Electropolished 3 μm
  • Weight
    15
  • Column i.d.
    100 mm
  • Tube height
    500 mm
DESCRIPTION
Glass chromatography columns for process development or biopharmaceutical manufacture, designed for hygienic operation and easy, efficient packing. Scalable from 100 mm to 450 mm inner diameter (i.d.) and bed volumes from 2 up to 121 liters. Pressure rating from 2.5 bar to 8 bar, depending on column size. Glass column tube allows efficient and reproducible packing. Material Certificates Reports (BPG MCR) available for ordering. Installation and Operational Qualification Documentation Packages available. Hygienic design and operation. Microbial attachment and growth is prevented through the use of calibrated precision glass, high-grade electropolished stainless steel and an absence of dead pockets. Easy, efficient packing and running with the single-screw adapter. All polymeric materials meet the requirements for USP class VI, described in USP Biological Reactivity Tests, in vivo. Isolating the column after packingSanitary stainless steel valves of the appropriate inner diameter are recommended to prevent contamination of the packed bed. The 2-way or 4-way valves with 6 mm internal diameter are suitable for BPG 100, 140, and 200 columns. The 10 mm internal diameter is suitable for BPG 300 and 450 columns. For storage purposes, the 25 mm blind flange with a clamp and gasket can be used to seal off the column.Connecting the column to your systemA clamp and gasket, 6 or 10 mm internal diameter, are required to connect the 25 mm sanitary flanged inlet/outlet to either valves or tubing of the same type. Preflanged tubing in 6 and 10 mm internal diameter is available.Assembly/disassembly of columnA torque wrench with an appropriate sized socket is required and can be ordered separately.What else do you need?Stand kits must be ordered separately.BPG 100 stand kit has adjustable feet; wheels with brakes are available: product code 18-1031-10. BPG 140 and 200 stands have wheels with brakes as standard: product code 18-1031-20. BPG 300 and 450 are supplied with stainless steel stand with wheels and foot-operated brakes.Useful spare parts and accessoriesNets: The column is delivered with 23 µm (polypropylene) nets. For resins with an average particle diameter