LXV325.99.30002
TSS HT sc Turbidity and Suspended Solids Inline Probe, Stainless Steel
6882
Add to Cart
Automatic Wiper
No
Cable Length
10 m (optional extension cables available)
Calibration Method
Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure.
Flow
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length
332 mm
Maintenance Interval
1 h/month
Material
Optics Carrier and Sleeve: stainless steel 1.4460 / stainless steel 1.4571
Max. cable length Sensor-to-Controller
100
Measuring Principle
Turbidity: 2-channel 90° scattered light measurement in accordance with DIN EN ISO 7027TSS: Modified absorption measurementEight-channel multiple-angle measurement, wavelength = 860 nm
Model
TSS HT sc
Operating Temperature Range
0 - 90 °C
Parameter
Turbidity, Suspended Solids
Pressure Range
< 10 bar or < 100 m
Range
Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L
Repeatability
TSS content: < 4 %Turbidity: < 3 %
Response Time T90
1 - 300 s adjustable
Special Notes
Installation note:Distance sensor–wall > 10 cm (TSS)
Warranty
12 months
Weight
approx. 1.6 kg
What's included?
Turbidity & Suspended Solids sensor, user manual
LXV328.99.30002
TSS EX1 sc Turbidity and Suspended Solids Inline Probe, Stainless Steel
7270
Add to Cart
Automatic Wiper
No
Cable Length
10 m (optional extension cables available)
Calibration Method
Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure.
Flow
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length
488 mm
Maintenance Interval
1 h/month
Material
Optics Carrier and Sleeve: stainless steel 1.4460 / stainless steel 1.4404
Max. cable length Sensor-to-Controller
100
Measuring Principle
Turbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO7027TSS: Modified absorption measurementEight-channel multiple-angle measurement, wavelength = 860 nm
Model
TSS EX1 sc
Operating Temperature Range
-10 - 50 °C
Parameter
Turbidity, Suspended Solids
Pressure Range
10 bar or 100 m
Range
Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L
Repeatability
TSS content: < 4 %Turbidity: < 3 %
Response Time T90
1 - 300 s adjustable
Special Notes
Installation note:Distance sensor–wall > 10 cm (TSS)
Warranty
12 months
Weight
approx. 2.7 kg
What's included?
Turbidity & Suspended Solids sensor, user manual
LXV327.99.10002
TSS XL sc Turbidity and Suspended Solids Inline Probe, Stainless Steel
6882
Add to Cart
Automatic Wiper
No
Cable Length
10 m (optional extension cables available)
Calibration Method
Turbidity (TRB): Factory calibratedSolids (TS): To be calibrated by customer on siteZero point: Permanently calibrated in the factory
Flow
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length
TSS XL sc: 232 mm
Maintenance Interval
1 h/month
Material
Gasket: FKM; Wiper: PA (GF), TPV
Measuring Principle
Combined multiple beam alternating light method with infrared diode system and beam focusingTurbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027, wavelength = 860 nmSolids: Modified absorption measurement; eight-channel multiple-angle measurement, wavelength = 860 nm
Model
TSS XL sc
Operating Temperature Range
0 - 80 °C
Parameter
Turbidity, Suspended Solids
Pressure Range
16 bar or 160 m
Range
Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L
Repeatability
TSS content: < 4 %Turbidity: < 3 %
Response Time T90
1 - 300 s adjustable
Special Notes
Installation note:Distance sensor–wall > 50 cm (Turbidity)> 10 cm (TSS)
Warranty
12 months
Weight
approx. 1.5 kg
What's included?
Turbidity & Suspended Solids sensor, user manual
LXV326.99.10002
TSS Vari sc Turbidity and Suspended Solids Inline probe, Stainless Steel
6882
Add to Cart
Automatic Wiper
No
Cable Length
10 m (optional extension cables available)
Calibration Method
Turbidity (TRB): Factory calibratedSolids (TS): To be calibrated by customer on siteZero point: Permanently calibrated in the factory
Flow
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length
TSS VARI sc 232 mm
Maintenance Interval
1 h/month
Material
Gasket: FKM; Wiper: PA (GF), TPV
Measuring Principle
Combined multiple beam alternating light method with infrared diode system and beam focusingTurbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027, wavelength = 860 nmSolids: Modified absorption measurement; eight-channel multiple-angle measurement, wavelength = 860 nm
Model
TSS Vari sc
Operating Temperature Range
0 - 80 °C
Parameter
Turbidity, Suspended Solids
Pressure Range
<16 bar or <160 m
Range
Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L
Repeatability
TSS content: < 4 %Turbidity: < 3 %
Response Time T90
1 - 300 s adjustable
Special Notes
Installation note:Distance sensor–wall > 50 cm (Turbidity)> 10 cm (TSS)
Weight
approx. 1.5 kg
What's included?
Turbidity & Suspended Solids sensor, user manual
LXV440.99.00002
AN-ISE sc Low cost ISE combination Ammonium and Nitrate probe (immersion) with RFID, 10 m cable
11091
Add to Cart
Cable Length
10 m (32.8 ft) fixed cable at sensor, available up to 100 m
Calibration Method
With Cartrical plus technology: automatic import of factory calibration data from cartridge to probe by RFID; 1 and 2 point matrix correction
Flow
< 4 m/s max. velocity
Length
320 mm
Material
Cartridge: stainless steel (1.4571), PVC, POM, ABS, NBRCleaning unit: TPE, PUR, stainless steel (1.4571)Sensor: stainless steel (1.4571), ASA + PC, silicon, PVC and PU
Operating Temperature Range
Air: -20 - 45 °C (-4 - 113 °F)
Parameter
NH4-N and NO3-N
Protection Class
IP68
Range
0 - 1000 mg/L NH4-N0 - 1000 mg/L NO3-N0 - 1000 mg/L K+0 - 1000 mg/L Cl-
Repeatability
± 5 % of the measured value (with standard solution)+ 0.2 mg/L NH4-N and NO3-N
Storage Conditions
5 to 40 °C (41 to 104 °F) sensor cartridge
Warranty
12 months
Weight
2.38 kg
What's included?
Ammonium & Nitrate probe, sensor cartridge, user manual
HACH9070800
USB Cable
28
Add to Cart
HACH3422A1A
Contacting Conductivity Sensor, Low Conductivity (k=0.05), with ½
682
Add to Cart
Cable Length
6 m (20 ft)
Flow
0 - 3 m/s maximum, fully immersed
Length
193.5 mm
Material
Compression Fitting: 1/2" NPT PVDF
Operating Temperature Range
-20 - 200 °C
Pressure Range
1.7 bar at 150 °C
Range
0 - 100 µS/cm
Repeatability
±0.5 % of reading
Response Time T90
30 s
Warranty
12 months
What's included?
Includes: sensor with 6 m (20 ft) cable and manual
HACH3433B8A
General Purpose Contacting Conductivity Sensor, PPS Body, Low Conductivity (k=0.5)
564
Add to Cart
Cable Length
20 ft
Flow
0 - 3 m/s
Length
134.6 mm
Operating Temperature Range
-20 - 200 °C
Pressure Range
6.8 bar at 150 °C (when used with hardware, a lower rated mounting hardware or piping material maylimit the temperatureand pressure ratings)
Range
0 - 1000 µS/cm
Repeatability
±0.5 % of reading
Response Time T90
< 30 s
Warranty
12 months
What's included?
Includes: sensor with 6 m (20 ft) cable and manual
HACH3422B3A
Contacting Conductivity Sensor for Low Conductivity, K=0.5, 3/4
682
Add to Cart
Cable Length
6 m (20 ft)
Flow
0 - 3 m/s maximum, fully immersed
Length
193.5 mm
Material
Compression Fitting: 3/4" NPT PVDF
Operating Temperature Range
-20 - 200 °C
Pressure Range
1.7 bar at 150 °C
Range
0 - 1000 µS/cm
Repeatability
±0.5 % of reading
Response Time T90
30 s
Warranty
12 months
What's included?
Includes: sensor with 6 m cable and manual
HACH3422C3A
Contacting Conductivity Sensor, Mid-Range Conductivity (k=1.0), with ¾
682
Add to Cart
Cable Length
6 m (20 ft)
Flow
0 - 3 m/s maximum, fully immersed
Length
193.5 mm
Material
Compression Fitting: ¾" NPT PVDF
Operating Temperature Range
-20 - 200 °C
Pressure Range
1.7 bar at 150 °C
Range
0 - 2000 µS/cm
Repeatability
±0.5 % of reading
Response Time T90
30 s
Warranty
12 months
What's included?
Includes: sensor with 6 m (20 ft) cable and manual