Altitude
3000 m (9842 ft) maximum
-
Analog Output Functional Mode
Linear, PID
-
Communication (optional)
Analog:Five 0-20 mA or 4-20 mA analog outputs on each analog output moduleUp to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.Digital:Profibus DPV1 moduleModbus TCPProfinet IO moduleEthernet IP module
-
Compatible Instruments
Compatible Sensors and Analyzers / Software Version (Release Year)Amtax sc / V2.30 (2018) or higherA-ISE sc / V1.02 or higherAN-ISE sc / V1.08 (2013) or higherN-ISE sc / V1.02 or higherNitratax clear sc, Nitratax eco sc, Nitratax plus sc / V3.13 (2013) or higherNT3100sc/NT3200scPhosphax sc / V2.30 (2018) or higherPhosphax sc LR/MR/HR / V1.01 (2018) or higherTSS sc / V41.73 (2013) or higherSolitax sc / V2.20 (2013) or higherTU5300sc, TU5400sc / V1.34 (2017) or higherSS7 sc (in Bypass) / V1.06 (2006) or higherUltraturb sc / V3.06 (2017) or higher1720E / V2.10 (2006) or higherSonatax sc / V1.15 (2016) or higherCL17sc / V2.7 (2019) or higherCL10sc / V1.14 (2013) or higher9184sc, 9185sc, 9187sc* / V2.03 (2013) or higherUvas plus sc / V3.01 (2017) or higherLDO 2 sc* / V1.22 (2013) or higher3798sc* / V2.03 (2013) or higher3700sc + Inductive Conductive Digital Gateway 6120800 / V3.00 (2017) or higher3422sc, Analog 3400 + Contacting Cond. Digital Gateway 6120700 / V3.00 or higherpHD sc*, pHD-S sc / V3.10 (2016) or higher1200-S sc* / V2.04 (2013) or higher pHD analog + Digital Gateway 6120500 / V3.00 (2017) or higherRC and PC analog sensor + Digital Gateway for conventional analog pH and ORP sensors 6120600 / V3.00 (2017) or higher 8362sc* / V3.00 (2017) or higher*Hardware Version1 of instrument is not supported
-
Compatible Network Technologies
GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.)CDMA (e.g. Verizon)
-
Compliance Certifications
CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C م (Morocco)
-
Conduit Openings
½" NPT conduit
-
Dimensions
½ DIN - 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.)
-
Display
3.5-inch TFT colour display with capacitive touchpad
-
Enclosure Rating
UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4XMetal enclosure with a corrosion-resistant finish
-
Indoor/Outdoor
Outdoor installation in direct sunlight or UV radiation requires UV protection screen and/or sunroof
-
Installation Category
Category II
-
Material Enclosures
Polycarbonate, aluminum (powder coated), stainless steel
-
Measurements
Two device digital SC connectors
-
Mounting
Wall, Pole, or Panel Mounting
-
Network Connectivity
LAN: Two Ethernet connectors (10/100 Mbps)Cellular: External 4GWi-Fi
-
Operating Temperature Range
-20 to 60 °C (-4 to 140 °F) (8 W (AC)/9 W (DC) sensor load)-20 to 45 °C (-4 to 113 °F) (28 W (AC)/20 W (DC) sensor load) Linear derating between 45 and 60 °C (-1.33 W/°C)
-
Outputs
LAN + Modbus TCP
-
Pollution Degree
4
-
Power Requirements (Voltage)
100-240 VAC ±10%, 50/60 Hz; 1 A
-
Power Supply
With EU plug
-
Protection Class
I, connected to protective earth
-
Relays
Two relays (SPDT);Wire gauge: 0.75 to 1.5 mm² (18 to 16 AWG)AC controllerMaximum switching voltage: 100 - 240 VACMaximum switching current: 5 A Resistive/1 A Pilot DutyMaximum switching power: 1200 VA Resistive/360 VA Pilot DutyDC controllerMaximum switching voltage: 30 VAC or 42 VDCMaximum switching current: 4 A Resistive/1 A Pilot DutyMaximum switching power: 125 W Resistive/28 W Pilot Duty
-
Sensor Input #1
Digital
-
Sensor Input #2
Digital
-
Software available
RTC-N/DNRTC-N/DN(ORP)RTC-N/DN(DO)
-
Storage Conditions
-20 - 70 °C (-4 - 158 °F), 0 - 95% relative humidity, non-condensing
-
USB Port
Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor.
-
Warranty
12 months
-
Weight
3.7 lb (controller only, w/o modules)
-
What's included?