DESCRIPTION
AcroPak™ 500, 1000, & 1500 capsules with Supor™ membrane, a Pall Life Sciences product, deliver fast, final filtration and high throughputs for liquid volumes up to 150 L
Provides higher throughputs and faster flow rates than similarly-sized competitive devices
Ideal for solutions requiring low protein binding
Saves money by increasing throughput with available built-in prefilter
100% integrity testing assures sterile filtrate
0.1 μm Supor membrane version reduces mycoplasma levels in cell culture media
Fusion-welded components eliminate the potential for release of extractables from sealing adhesives
Sterilization by gamma irradiation eliminates potential contamination by EtO residuals
Materials of Construction
Filter Media: Supor [hydrophilic polyethersulfone (PES)] membrane
Housing, Vent Plug, and Filter Support: Polypropylene
Sealing Technology: Thermal bonding
Filling Bell: Polycarbonate
Pore Size
0.1/0.1, 0.2/0.2, 0.8/0.2, and 0.8/0.45 μm
Effective Filtration Area
500 Series: 500 cm2
1000 Series: 1000 cm2
1500 Series: 1500 cm2Dimensions
Housing Diameter: 6.9 cm (2.7 in.)
Overall Length (Without Filling Bell):
500 Series: 14.5 cm (5.7 in.)
1000 Series: 19.3 cm (7.6 in.)
1500 Series: 22.9 cm (9.0 in.)
Inlet/Outlet Connections
500 and 1000 Series: 6.4 - 12.7 mm (1/4 - 1/2 in.) stepped hose barb with filling bell on outlet
1500 Series: 12.7 mm (1/2 in.) straight hose barb, no filling bell on outlet
Typical Hold-Up Volume (with 10 psi air purge)
500 Series: 30 mL
1000 Series: 45 mL
Maximum Operating Temperature
60°C (140°F) at 2.1 bar (210 kPa, 30 psi)
Maximum Operating Pressure
Continuous: 4.1 bar (410 kPa, 60 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point – Water
0.2 μm: 3.5 bar (350 kPa, 51 psi)
0.45 μm: 1.7 bar (170 kPa, 24 psi)
60% IPA/40% h3O (v:v)
0.1 μm: 2.4 bar (240 kPa, 35 psi)
Bacterial Retention
0.1/0.1 μm Capsules: Lot samples retain a minimum of 107 cfu/cm2 of A. laidlawii challenge per modified ASTM F838, current revision
0.2 μm Capsules: Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level
Biological Safety
Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo
Sterilization
Sterilized by gamma irradiation
If desired, autoclave once only prior to use at 121 - 123°C (250 - 253°F) for a maximum of 20 min.
Applications
Large-volume media preparation
Pilot-scale manufacturing
Liquids requiring prefiltration, such as serum-containing media
Sterile filtration of buffers
Point-of-use filtration for laboratory water
Reduce mycoplasma with the 0.1 µm pore size