Bacterial Endotoxins
A sample aqueous extraction contains less than 0.5 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.
-
Bacterial Retention
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
-
Catalogue Number
MPGL10CB1
-
Chemistry
Polyvinylidene Fluoride (PVDF)
-
Device Configuration
Capsule
-
Diameter
1/4 in.
-
Good Manufacturing Practices
These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
-
Gravimetric Extractables
The extractables level was equal to or less than 2.5 mg per device after 24 hours in water at controlled room temperature.
-
Inlet Connection
NPT
-
Inlet Connection Diameter
1/4 in.
-
Length
13 cm (5.12in.)
-
Material Size
1
-
Media
Durapore®
-
Non-Fiber Releasing
This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
-
Outlet Connection
NPT
-
Outlet Connection Diameter
1/4 in.
-
Pore Size
0.22 µm
-
Sterility
Other
-
Sterilization
3 autoclave cycles of 90 min @ 123°C; not in-line steam sterilizable
-
Trade Name
Millipak®
-
USP Bacterial Endotoxins
A sample aqueous extraction contains less than 0.5 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.
-
Wettability
Hydrophilic