Post autoclave (1 cycle of 30 minutes at 123 °C) and pure water flush of 100 L/m2 (10 L/ft2) of media surface area.NVR: 1.50 – 2.02 mg/m2TOC: 416– 593 ppb
Material Size
2
Maximum Differential Pressure, bar (psid)
Forward: 2.07 bar (30 psid) @ 25 °C; 1.03 bar (15 psid) @ 80 °C; 0.21 bar (3 psid) @ 123 °C. Reverse: 0.03 bar (0.5 psid)
Maximum Inlet Pressure, bar (psi)
2.1 bar (30 psi) @ 25 °C; 1 bar (15 psi) @ 80 °C; 0.2 bar (3 psi) @ 123 °C
Oxidizable Substances
The effluent of an autoclaved (1 cycle of 30 minutes @ 123 °C) stacked device is negative after a WFI flush of 10 liters per ft2 (100 L/m2) of media surface area
Sanitization
May be hot water sanitized for 20 cycles using purified water @ 80 °C for 30 minutes
Seal Material
Silicone (SI)
Sterilization
May be steam sterilized for 3 steam-in-place (SIP) cycles of 30 minutes @ 123 °C
Trade Name
Millistak+®
Wettability
Hydrophilic
DESCRIPTION
Post-centrifuge or settled permeate containing cellular particulate